We have collected the most relevant information on Audiometry Masking Rules. Open the URLs, which are collected below, and you will find all the info you are interested in.


Audiometric Masking

    https://www.interacoustics.com/guides/test/audiometry-tests/audiometric-masking#:~:text=1%20Perform%20air%20conduction%20audiometry%20unmasked%20for%20both,Select%20NB%20in%20channel%202.%20More%20items...%20
    none

MASKING IN PURE TONE AUDIOMETRY - ENT Lectures

    http://entlectures.com/Resources/Dialogs/Dialogue%20with%20the%20ENT%20Residents_Masking%20in%20pure%20tone%20audiometry.pdf
    Rule 1: Masking is needed at any frequency where the difference between the left and right not-masked A/C thresholds is 40 dB or more when using supra- or circum-aural earphones or 55 dB or when using insert earphones

Audiometric Masking - Interacoustics

    https://www.interacoustics.com/guides/test/audiometry-tests/audiometric-masking
    Channel 1 should be set to the left ear (test ear) using the preferred stimulus (usually tone). The masking frequency will automatically change along with the tone frequency when masking is turned on. You can set the masking and tone frequencies by …

How to Perform Masking in Audiology - eMoyo

    https://blog.emoyo.tech/content/masking-in-audiology
    The formula for deciding whether you should mask or not: Mask if the air-bone gap in the test ear ≥ 10 dB. How much masking is needed? To mask the non-test ear for for bone-conduction testing, add 10 dB HL to the air-conduction threshold in the non-test ear and add the occlusion effect (air conduction non-test ear + 10 dB HL + occlusion effect).

Clinical Masking for Audiometric Testing in Adults - …

    http://cshbc.ca/wp-content/uploads/2019/03/CSHBC-PROT-QA-03-Clinical-Masking-for-Audiometric-Testing-in-Adults.pdf
    1. Select initial amount of masking for the non-test ear (i.e., air-conducted threshold of the non-test ear (AC Nontest + 10 dB HL), plus correction factor for occlusion effect, if applicable ; 2. Re-establish threshold in the test ear with this initial amount of masking (e.g., AC Nontest Ear+10); 3.

Speech audiometry & masking - SlideShare

    https://www.slideshare.net/bethfernandezaud/speech-audiometry-masking
    Just like masking for puretones Determine the need for masking and use plateau method to determine effectively masked threshold for speech Consider the interaural attenuation (IA): 40 dB for headphones 70 dB for inserts 0 dB for bone (you probably won’t do a bone-conducted speech test, though it is possible and very useful in children) The same rules apply …

Guidelines for Manual Pure-Tone Threshold Audiometry

    https://www.asha.org/policy/GL2005-00014/
    Masking for diagnostic audiometry. Appropriate masking should be applied to the nontest ear when the air-conduction threshold obtained in the test ear exceeds the interaural attenuation to the nontest ear. Because the procedures for masking are not confined to pure-tone measures, these procedures are not discussed in this set of guidelines.

Masking Rules for Hearing Care Professionals

    https://hearinghealthmatters.org/hearinginternational/2020/masking-rules-for-hearing-care-professionals/
    Stevens and Davis (1938) said, “There are, in general, two conditions under which a normally effective auditory stimulus may fail to arouse a sensation. One is when it is accompanied by another sound which obliterates or masks it; the other is when it is preceded by a sound which leaves the organism unresponsive or fatigued.”.

Now you know Audiometry Masking Rules

Now that you know Audiometry Masking Rules, we suggest that you familiarize yourself with information on similar questions.